CO2 Gas Calibration with eVol®, the digital analytical syringe

Quantification of CO2 Capture in Aqueous Solutions of Alcanolamine

Alain Valtz Dr. Christophe Coquelet, CEP (Center of Energy and Processes) Mines ParisTech, France

MINES ParisTech is one of France’s oldest higher education institutions. While its inspirational model has not changed over the years, its teaching methods and research have been at the leading edge since 1783. Two Nobel laureates trained at the School: Georges Charpak (Physics, 1992) and Maurice Allais (Economics, 1988).

MINES ParisTech is established in five major fields: Earth sciences and the environment; Energy and process engineering; Mechanical engineering and materials; Mathematics and systems; Economics, management and society.

The Center of Energy and Processes (CEP) has skills useful for studying the transformation of matter and energy in many fields, focussing on complex energy systems, particularly in transient conditions, and on controlling their emissions.

The TEP laboratory specializes in the measurement of thermodynamic properties. This laboratory, “Thermodynamique des équilibres entre phases”, provides industry with specialists skilled in both experimental and theoretical aspects of handling fundamental and applied works successfully.

Figure 1. Magnifying tool developed to improve the accuracy, precision and consistency of syringe use.

Description

A key area of study for the TEP lab is the measurement of partition coefficients (ELV) of mercaptans for aqueous amines in the presence of carbon dioxide and/or hydrogen sulfide. The release of CO2 is measured using a GC method with a TCD detector. A calibration regime is essential because of the quantitative method of analysis. Samples may be either liquid or gaseous. Calibrations are required before, during and after the testing process which takes place over several months. The accuracy and precision of the calibrations are critical.

The GC system is calibrated by injecting a known volume of CO2 aspirated directly from a gas bottle through a septum. This is done with manual gas tight syringes and is operator dependent, often resulting in inaccurate measurements.

To reduce discrepancies a magnifying tool was developed to improve the accuracy, precision and consistency of syringe use by different operators. Even with the new tool the calibration process remains time consuming and operator dependent. To address these methodology issues, CEP/TEP lab have evaluated and now use eVol®, the world’s first digital analytical syringe.

Method

eVol® is used to collect CO2 from a gas bottle through a regulator equipped with a septum port. Using the 50 μL and 500 μL eVol® XCHANGE® syringes, volumes of gas are aspirated and injected into the GC system injection port.

The CO2 concentration is directly proportional to the sample volume. It elutes on a packed column in less than 50 seconds with levels detected by a TCD detector.

Each time the total aspirated volume is injected. Increasing volumes of aspirated sample are successively injected to set up the calibration curve with the test for each volume repeated several times.

Tests are performed in a stable and controlled environment at 23 °C and an atmospheric pressure of 1006.3 mbar.

RESULTS

Figure 2. Linear regression of the calibration curve, R2= 0.9991

Calibration using eVol® 50 μL syringe

Injection of 64 aspirated volumes from 2 μL to 50 μL.

Figure 3. Accuracy of several injected volumes from 2 μL to 50 μL.

Figure 4. Linear regression of the calibration curve R2= 0.9999.

Calibration using the 500 μL syringe

Injection of 82 aspirated volumes from 20 μL to 500 μL.

Figure 5. Accuracy of several injected volumes from 20 μL to 500 μL.

Discussion

The accuracy of low volumes (2.0 μL and 2.5 μL) using the eVol® 50 μL syringe is affected by Helium back pressure, which causes a dilution of CO2 during the injection process. The dispersion observed for low volumes using the 50 μL syringe and reported in figure 2 is not observed in figure 4 when the 500 μL syringe is used.

Conclusions

Using eVol® for injection of different volumes of CO2, CEP/TEP was able to achieve calibration of their instrument with levels of accuracy and precision not previously obtained using manual syringes.
In addition, the needles on the eVol® syringes allow for sample collection directly from the gas bottle – similarly to the manual syringes used previously.
Using the eVol®, CEP/TEP lab increased the accuracy of their calibration while reducing the handling time involved in this process.

Facts, background information, dossiers
  • needles
  • matter
  • hydrogen
  • helium
  • emissions
  • concentration
  • carbon dioxide
  • amines
  • SGE
More about SGE
  • White papers

    CO2 Gas Calibration with eVol®, the digital analytical syringe

    Using eVol® for injection of different volumes of CO2, CEP/TEP was able to achieve calibration of their instrument with levels of accuracy and precision not previously obtained using manual syringes. In addition, the needles on the eVol® syringes all more

    GC-C-IRMS Determination of Hormones

    By amending methodology to utilize eVol® for sample preparation processes for GC-C-IRMS C isotope ratio analysis, the ISORA department has: improved sample preparation conditions and the GC-C-IRMS analysis results by removing the risk of ghost peaks, more

  • News

    Trajan completes acquisition of SGE business

    Trajan Scientific and Medical, “Trajan” has completed its acquisition of the Chromatography and Liquid Handling businesses of SGE Analytical Science, “SGE”. SGE was founded by Ern Dawes in Melbourne Australia in 1961 and has remained in the Dawes family for over 50 years. Its business, focu ... more

    SGE Analytical Science: eVol Wins prestigious R&D100 Technology Innovation Award

    eVol®, the world’s first automated analytical syringe created and manufactured by SGE Analytical Science, has been selected as a 2010 R&D100 Award winner. eVol® is a unique laboratory tool that removes the human limitation of manual liquid handling delivering significantly improved levels ... more

    The world´s first hand-held automated analytical syringe by SGE is nominated for the GIT-InnovationsAward 2010

    eVol®, a revolution in liquid handling of smallest volumes is nominated for the GIT-InnovationsAward 2010. An independent jury of renown university professors from Germany - has made a pre-selection from the huge number of submitted products and nominated eVol® for the honouring. The aim ... more

  • Companies

    SGE Analytical Science Europe Ltd

    SGE is a manufacturer of accessories and ancillary equipment for chromatography, biotechnology and analytical instrumentation. SGE has been in the analytical (chromatography) business for over 40 years and today they are world leaders in areas such as microsyringes and Capillary Column tech ... more

Your browser is not current. Microsoft Internet Explorer 6.0 does not support some functions on Chemie.DE