Detection of Avian H5N1 Virus with the LightCycler® Instrument

Cornelia Götz¹, Ulrich Lass¹, Roger Petersen¹, Massimo Ulivi², and Olfert Landt¹*
¹TIB MOLBIOL GmbH, Berlin, Germany
²TIB MOLBIOL srl., Genova, Italy

*Corresponding author

Introduction

The avian H5N1 influenza virus has killed thousands of birds and caused several fatal cases in humans in Asia and Turkey. A fast and reliable test to identify the virus is urgently required for an effective epedemiological control, and associated medical research activities.

RNA viruses mutate easily. Thus, detection has to focus on conserved sequence regions. The preferred gene for the detection of a broad spectrum of influenza A virus variants is the matrix protein (M2) gene. Tests for the specific detection of the H5N1 virus must detect its hemagglutinin (H5) and neuraminidase (N1) genes.

Real-time PCR-based RNA virus tests using FRET hybridization probes are the method of choice. They also detect newly occurring variants as a specific signal in the melting curve analysis, even when a few bases at the binding site are exchanged.

Materials and Methods

To develop influenza A specific real-time PCR assays, a cDNA sample obtained from Thailand and derived cloned fragments were used. Tests were verified with a patient-derived sample from Hong Kong.

Using the Transcriptor First Strand cDNA Synthesis Kit, 10 µl of RNA were incubated for 30 minutes at room temperature to generate cDNA. 5 µl of cDNA were amplified using the LightCycler® FastStart DNA MasterPLUS HybProbe. A one-step RT-PCR protocol is available from the authors (currently in the process of optimization).

Results

We first demonstrated the accuracy and efficiency of the premixed LightMix® real-time PCR assay for H5 using the LightCycler® 1.2 Instrument during a conference held in Khon Kaen, Thailand, in October 2004.

Coincidentally, the PCR fragment amplified with this test was part of the amplicon recommended by the WHO in June 2005. The location of the sequence mutations in the receptor-binding site of the H5 gene, recently reported in one isolate from Turkey, is included in this PCR fragment but does not overlap with primer or probe binding sites.

To duplex the H5 assay with an N1 specific target and achieve a sensitivity of at least ten genome copies (plasmid) for each targeted gene, we had to improve the H5 primers. The newly developed duplex assay was run with the LightCycler® 2.0 and LightCycler® 480 Instruments. In both cases the sensitivity was within ten copies (Figure 1).

The duplex H5N1 PCR was complemented with a PCR assay for the Matrix protein gene (M2) using the LightCycler® 2.0 Instrument, allowing detection of a broad range of influenza A virus types. We combined the probes from our earlier design [1] with published primers from the German Reference Laboratory [2] and adapted the primers to detect all new H5N1 isolates. The M2 PCR was completed by an Internal Control PCR based on the amplification of a fragment from Lambda DNA to verify the absence of PCR inhibitors.

All established primer systems have been combined with hydrolysis probes, sharing the binding sites of the HybProbe probes, yielding similar results in our experiments. In addition, we submitted some of the published H5 specific real-time PCR assays to a comparative test. In particular, the WHO-recommended primers derived from Yuen et al. [3, 4] performed well, whereas the hydrolysis probe assay from Spackman et al. [5] did not perform as expected. This was due to mismatches for the H5N1 target found in the forward primer H5+1456. Neither did the triplex hydrolysis probe assay from Payungporn et al. [6] give satisfying results in single PCR runs, and the triplex assay performed using the LightCycler® 480 Instrument was evaluated as inadequate. However, in contrast, both H5-­specific hydrolysis probe assays from Ng et al. [7] (US CDC recommendation) showed a superior performance, reaching a sensitivity of fewer than ten genome copies in single-assay runs with the LightCycler® 2.0 Instrument and LightCycler® 480 Instrument (Figure 2).

References

1. Smith AB (2003) J Clin Virol 28: 51–58

2. Schweiger B (2000) JCM 38: 1552–1558

3. Recommended laboratory tests to identify avian influenza A virus in specimens from humans (2005) WHO Geneva

4. Yuen KY et al. (1998) Lancet 351: 467–471

5. Spackman E et al. (2003) JCM 40: 3256–3260

6. Payungporn S et al. (2005) Virol Methods

7. Ng EKO et al. (2005) Emerg Infect Dis

 

LightMix® products consist of premixed primers and hybridization probes for use with LightCycler® Instrument series. The HybProbe probe is lyophilized in 6 individual vials, 16 reactions each, providing a total of 96 reactions. Also included is a lyophilized standard row for quantitative analysis ranging from 10 to 106 target equivalents per reaction. The product does not contain Taq DNA polymerase or dNTPs. The product is designed for life science research only.

The LightMix® InfA kits have been tested by the German Reference Laboratories for Human Influenza (Robert-Koch-Institut) and for Veterinarian Influenza (Friedrich-Löffler-Institut) to be ‘suitable for the detection of the AIV genome’ and to detect all current H5N1 isolates.

 

This article was originally published in Biochemica 2/2006, pages 12-13. ©Springer Medizin Verlag 2006