Abstract

Monitoring serum/plasma concentrations of lamotrigine may be useful under certain circumstances. An HPLC column packed with strong cation‐exchange (SCX)‐modified microparticulate silica together with a 100% methanol eluent containing an ionic modifier permits direct injection of sample extracts. An HPLC‐UV method developed using this principle for the measurement of serum/plasma lamotrigine is simple, sensitive and selective. The analysis time is less than 5 min. Intra‐ and inter‐assay precision and accuracy meet acceptance criteria, and sample stability, and potential interferences from other compounds have been evaluated. There was good agreement with consensus mean results from external quality assessment samples (n = 32). Analysis of patient samples (n = 115) using the HPLC method and the Seradyn QMS® Lamotrigine immunoassay showed that the immunoassay over‐estimated lamotrigine by 21% on average. Copyright © 2010 John Wiley & Sons, Ltd.

Autoren:   Morgan, Phillip E.; Fisher, Danielle S.; Evers, Richard; Flanagan, Robert J.
Journal:   Biomedical Chromatography
Band:   25
Ausgabe:   7
Jahrgang:   2011
Seiten:   775
DOI:   10.1002/bmc.1515
Erscheinungsdatum:   01.07.2011